A range of high performance, yet compact test chambers. The low operating noise levels make them ideal for laboratory use and they offer good test space capacity with a small footprint making them useful for any application where space is limited. They are also highly mobile and only require a single phase electrical supply.
Standard Features:
- Available in capacities of 34, 64 or 100 litres
- Temperature Range: +10°C to 180°C, -40°C to +180°C or -70°C to +180°C
- Humidity Range: 10%RH to 98%RH (climatic range +10°C to +95°C)
- Large observation window
- User-friendly
- Colour touchscreen for programming
- Interface Ethernet
- Interface USB for automatical measurement-documentation per
USB-stick
- Remote control and remote monitoring via intranet or internet
- Safety device for test specimens with independent temperature
limiters tmin / tmax
- Cable port
- 1 x shelf
|
|
|
| |
Options include:
- Temperature measuring on test specimen
- Capacitive humidity sensor
- Additional entry ports and shelves
- Programming and logging software
|
|
|
|
| |
| Voetsch Model VCL |
VCL0003 |
VCL4003 |
VCL7003 |
VCL0006 |
VCL4006 |
VCL7006 |
VCL0010 |
VCL4010 |
VCL7010 |
| Weiss Model WKL |
WKL34/+10 |
WKL34/40 |
WKL34/70 |
WKL64/+10 |
WKL64/40 |
WKL64/70 |
WKL100/+10 |
WKL100/40 |
WKL100/70 |
| Volume (Litres) |
34 |
34 |
34 |
64 |
64 |
64 |
100 |
100 |
100 |
| Internal chamber dimensions |
| Width mm |
350 |
350 |
350 |
470 |
470 |
470 |
490 |
490 |
490 |
| Depth mm |
300 |
300 |
300 |
345 |
345 |
345 |
400 |
400 |
400 |
| Height mm |
310 |
310 |
310 |
400 |
400 |
400 |
500 |
500 |
500 |
| External chamber dimensions |
| Width mm |
640 |
640 |
640 |
760 |
760 |
760 |
780 |
780 |
780 |
| Depth mm |
930 |
930 |
750 |
980 |
980 |
800 |
1105 |
1105 |
1105 |
| Height mm |
980 |
980 |
1730 |
1070 |
1070 |
1780 |
1190 |
1190 |
1190 |
| Chamber weight (kg) |
110 |
110 |
130 |
120 |
120 |
140 |
170 |
190 |
210 |
| Temperature Range |
| Max temp °C |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
| Min temp °C |
+10 |
-40 |
-70 |
+10 |
-40 |
-70 |
+10 |
-40 |
-70 |
Heat compensation (Watts max) |
-- |
800 |
550 |
-- |
800 |
550 |
-- |
1100 |
700 |
Rate of cooling (°C/min) with an empty chamber in accordance with IEC600068 |
2 |
4 |
4 |
3 |
3.5 |
3.5 |
2 |
3.5 |
3.5 |
Rate of heating (°C/min) with an empty chamber in accordance with IEC600068 |
3 |
6 |
3 |
3 |
5 |
2.5 |
3 |
5 |
3.5 |
Humidity Range between +10°C and +95°C |
10% RH to 98% RH |
| Electrical connection |
1/N/PE AC 230 V ± 10%, 50Hz, 16A |
|
|
|
|