A range of high performance, yet compact test chambers. The low operating noise levels make them ideal for laboratory use and they offer good test space capacity with a small footprint making them useful for any application where space is limited. They are also highly mobile and only require a single phase electrical supply.

Standard Features:

  • Available in capacities of 34, 64 or 100 litres
  • Temperature Range: +10°C to 180°C, -40°C to +180°C or -70°C to +180°C
  • Humidity Range: 10%RH to 98%RH (climatic range +10°C to +95°C)
  • Large observation window
  • User-friendly
  • Colour touchscreen for programming
  • Interface Ethernet
  • Interface USB for automatical measurement-documentation per USB-stick
  • Remote control and remote monitoring via intranet or internet
  • Safety device for test specimens with independent temperature limiters tmin / tmax
  • Cable port
  • 1 x shelf

   
 

Options include:

  • Temperature measuring on test specimen
  • Capacitive humidity sensor
  • Additional entry ports and shelves
  • Programming and logging software

     
 
 Voetsch Model VCL VCL0003 VCL4003 VCL7003 VCL0006 VCL4006 VCL7006 VCL0010 VCL4010 VCL7010
 Weiss Model WKL WKL34/+10 WKL34/40 WKL34/70 WKL64/+10 WKL64/40 WKL64/70 WKL100/+10 WKL100/40 WKL100/70
 Volume (Litres) 34 34 34 64 64 64 100 100 100
 Internal chamber dimensions
 Width mm 350 350 350 470 470 470 490 490 490
 Depth mm 300 300 300 345 345 345 400 400 400
 Height mm 310 310 310 400 400 400 500 500 500
 External chamber dimensions
 Width mm 640 640 640 760 760 760 780 780 780
 Depth mm 930 930 750 980 980 800 1105 1105 1105
 Height mm 980 980 1730 1070 1070 1780 1190 1190 1190
 Chamber weight (kg) 110 110 130 120 120 140 170 190 210
 Temperature Range
 Max temp °C +180 +180 +180 +180 +180 +180 +180 +180 +180
 Min temp °C +10 -40 -70 +10 -40 -70 +10 -40 -70
 Heat compensation
 (Watts max)
-- 800 550 -- 800 550 -- 1100 700
 Rate of cooling (°C/min) with
 an empty chamber in
 accordance with IEC600068
2 4 4 3 3.5 3.5 2 3.5 3.5
 Rate of heating (°C/min) with
 an empty chamber in
 accordance with IEC600068
3 6 3 3 5 2.5 3 5 3.5
 Humidity Range between
 +10°C and +95°C
10% RH to 98% RH
 Electrical connection 1/N/PE AC 230 V ± 10%, 50Hz, 16A
   
 