Home The Company Products Software Service & Support Hire Contact Us News Downloads


Our Product leaflets are also produced as PDFs.
Click below to download Adobe Reader.

download acrobat reader

WVC range of temperature humidity chambers, temperature humidity test, economical temperature humidity chamber

Environmental test chambers, temperature humidity test chambers, environmental stress screening


 

WVC Climatic Test Chambers

(back to main page)
 

   
   

A range of robust and reliable temperature and humidity test chambers with many features, which are usually optional but included as standard.

Standard features:

  • Available in a range of capacities from 190 litres to 1540 litres
  • Temperature Range: -70°C or -40°C to +180°C
  • Humidity Range: 10%RH to 98%RH (climatic range +10°C to +95°C)
  • Vapour tight welded stainless steel inner test space
  • Colour touchscreen for programming
  • Extensive networking capabilities via TCP/IP
  • Specimen protection against condensation
  • Multi-glazed heated window
  • Internal test space illumination
  • 1 x wire mesh shelf
  • 2 x cable access ports
  • 4 potential-free digital switches
  • Safety device for test specimens with independent temperature limiters tmin / tmax

   
 

Options include:

  • Additional shelves
  • Additional ports
  • Compressed air dryer
  • Programming and logging software

 Model WVC C180-40 C340-40 C340-70 C600-40 C600-70 C1000-40 C1000-70 C1500-40 C1500-70
 Volume (Litres) 190 335 335 600 600 990 990 1540 1540
 Internal chamber dimensions
 Width mm 580 580 580 800 800 1100 1100 1100 1100
 Depth mm 450 765 765 800 800 950 950 1475 1475
 Height mm 750 750 750 950 950 950 950 950 950
 External chamber dimensions
 Width mm 870 870 870 1090 1090 1390 1390 1390 1390
 Depth mm 1475 1790 1790 1845 1845 1995 1995 2520 2520
 Height mm 1780 1780 1780 1980 1980 1980 1980 1980 1980
 Chamber weight (kg) 425 475 520 610 675 850 950 930 1110
 Temperature Range
 Max temp °C +180 +180 +180 +180 +180 +180 +180 +180 +180
 Min temp °C -40 -40 -70 -40 -70 -40 -70 -40 -70
 Rate of cooling (°C/min) with
 an empty chamber in
 accordance with IEC600068
3.5 3.5 3 3 2.5 3 2.5 2.5 2.3
 Rate of heating (°C/min) with
 an empty chamber in
 accordance with IEC600068
4 3.2 3 4 4 4 4 3.5 3.5
 Humidity Range between
 +10°C and +95°C
10% RH to 98% RH
 Electrical connection 3/N/PE AC 400 V ± 10%, 50Hz
 Max. Nominal Current Amps 12 12 12 15 19 22 29 22 29


   
 

Please contact us or download the WVC leaflet for further details.


WVC leaflet download

   
 


T: +44(0)1495 305 555
F: +44(0)1495 303 595
E: sales@designenvironmental.co.uk