| MODELS: |
190- 40 |
190- 70 |
335- 40 |
335- 70 |
610- 40 |
610- 70 |
990- 40 |
990- 70 |
1550- 40 |
1550- 70 |
| Test volumes (litres) |
190 |
190 |
335 |
335 |
610 |
610 |
990 |
990 |
1550 |
1550 |
| Internal chamber dimensions |
| Height h mm |
750 |
750 |
750 |
750 |
950 |
950 |
950 |
950 |
950 |
950 |
| Width w mm |
540 |
540 |
540 |
540 |
750 |
750 |
1050 |
1050 |
1050 |
1050 |
| Depth d mm |
450 |
450 |
765 |
765 |
800 |
800 |
950 |
950 |
1450 |
1450 |
| External chamber dimensions |
| Height h mm |
1780 |
1780 |
1780 |
1780 |
1980 |
1980 |
1980 |
1980 |
1980 |
1980 |
| Width w mm |
860 |
860 |
860 |
860 |
1080 |
1080 |
1380 |
1380 |
1380 |
1380 |
| Depth d mm |
1300 |
1300 |
1630 |
1630 |
1690 |
1690 |
1880 |
1880 |
2330 |
2330 |
| TEMPERATURE RANGE DELTA H SERIES |
| Maximum temperature level in ºC |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
+180 |
| Minimum temperature level in ºC |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
-40 |
-70 |
| Rate of cooling in ºC/Min average |
[empty chamber] approx. 1.5ºC
to 2ºC/min |
| Rate of heating in ºC/Min average |
[empty chamber] approx. 1.5ºC
to 2ºC/min |
| Temperature fluctuation |
[measured at the control sensor after
stabilisation] approx. ±0.3ºC to ±0.5ºC |
| Temperature variation |
across chamber [following chamber stabilisation]
approx. ±1ºC to ±2ºC |
| TEMPERATURE RANGE WITH HUMIDITY OPERATIONAL
|
| Maximum temperature level in ºC |
+95 |
+95 |
+95 |
+95 |
+95 |
+95 |
+95 |
+95 |
+95 |
+95 |
| Minimum temperature level in ºC |
+10 |
+10 |
+10 |
+10 |
+10 |
+10+ |
+10 |
+10 |
+10 |
+10 |
| Humidity range |
from 10% to 98% RH |